4. Operating procedures
List of operation modes and functions
Functions of the unit are as shown below:
Name Description Page
1
Overheat prevention
function
Automatic overheat prevention function:
This function is linked to the unit set temperature and has
been set to so that it is automatically activated (returned
automatically) at a temperature 12 higher than the set
temperature in the bath.
Standalone overheat protection device (Thermostat)
When the temperature in the bath reaches the set
temperature of the standalone overheat protection device
(Thermostat), its heater circuit trips to shut off controller
operation.
The temperature can be set with the manual dial on the
hydraulic overheat prevention device installed at the right
side of the unit.
P. 1 5
2
Calibration
Offset function
Calibration offset function compensates any differences
between the target temperature in the bath and the control
temperature of the controller (sensor temperature.)
The function can compensate to either plus or minus side
for the whole temperature band of the unit.
This compensation can be set with the SUB MENU keys.
P. 2 3
3 Setting lock function
This function locks the set operation status.
The lock can be set or released with the SUB MENU key.
P. 2 4
4
Power outage
compensation function
This function returns the main unit operation to the resume
status after recovery from power outage, or keeps the
current stop status.
This compensation can be set with the SUB MENU keys.
P. 2 5